CS-400 Auto-Chemistry Analyzer
1. General characteristics
- Spectrophotometer of panoramic technical concave plane wave grid and beam optical path, helps to identify the microscopic volume of the reaction solution.
- Polished 60nm probe with liquid level detection, blockage and collision avoidance functions.
- The colorimetric cuvette automatically washes and detects empty cuvettes.
- Rerun sample and dilution function.
- The software operates in English and supports printing reports in many formats.
- Users can manually enter patient information in combined batches, suitable for quick setup for special hospitals, blood stations and health check-ups.
2. Mechanical system for sample / chemical aspiration:
- 3 separate needles, one for samples, two for chemicals. The suction needles are polished using nanotechnology, minimizing cross-contamination.
- Automatic liquid level detection: The needle automatically detects the liquid level, ensuring the needle enters the liquid at the perfect depth and minimizing contamination on the needle surface.
- 3 separate syringes, one for samples, two for chemicals. Syringe is made of long-life ceramic piston, ensuring accurate sample collection and low maintenance costs.
- 3 suction needles have anti-collision function, self-resetting, and automatic suction of samples and chemicals.
- Smart clog detection: the pump line has a pressure sensor that can detect clogs. The pump line has a high-pressure washing function to clear clogs, ensuring accurate suction volume.
3. Mechanical system of double stirring rod
- 2 separate stirring rods start stirring immediately after the machine adds chemicals to mix the reaction solution evenly.
- The surface of the stick is coated with telfon to avoid the formation of suspensions in the liquid and reduce cross-contamination.
- The stirrer has a paddle design and is washed in a rolling manner for the best cleaning effect
4. The reaction cuvette is washed automatically
- 7 stations with 11 steps, washed with warm water, ensuring complete cleanliness.
- Vacuum drainage, washing with warm water and specialized detergent help enhance cleaning ability.
5. QC and calibration function
- Linear and non-linear calibration, set with time, reagent lot number... with flexible calibration method.
- 9 types of calibration curves meet a variety of analytical requirements.
- Six different calibration levels can be used for each test.
- Has a calibration query function, plotting the trend of the calibrated K value, helping to minimize system errors.
- QC on demand and monthly QC, ensuring the machine operates in the best conditions.
- QC rules: default to Westgard rules
- QC chart: automatically draw and print QC, L-J and Cumulative charts.
- Automatically stores causes of machine out of control and primary data, helping to manage quality in the QC lab.